Timer Setup Loop

(Digital Plus Only)
Timer
To set the timer, with the bath on, press and hold the Timer button for five
seconds to access the controller's Timer Setup Loop. The display will indicate
diSP. Use the arrow buttons to select the desired clock display. Press the
Timer button and the controller will display cloc. Use the arrow buttons to
display the correct time.
Press the Timer button and the controller will display t-On. Press the arrow
buttons to set the desired turn-on time. Press the Timer button and the
controller will display t-OFF. Press the arrow buttons to set the desired turn-off
time. Pressing the Timer button after entering the clock time will display StorE.
Press YES to accept the values, NO to abort them.
If the Timer button is not pressed for 60 seconds the function will time-out and
the changes will not be saved. Pressing the Mute will also exit the Timer Setup
Loop and return you to the temperature display.
If the bath is turned off using
will go out and the timer mode will be disabled. With the bath off, you have to
press the Timer button to enable the timer mode and light the LED.
NOTE: Enabling serial communications disables the
TIMER mode. If an error condition exists: serial
communications enabled; time-on not defined; time-off
not defined; or clock not set, the unit will beep and
will not enable the function. The display will then
indicate the reservoir fluid temperature.
- 25 -
with the timer mode enabled, the Timer LED
If a power disruption occurs, the timer mode is
disabled. The Timer Setup Loop will have to be used
to restore the settings.