Modbus Protocol
Group Name
Setup/Relay 2/Control
Setup/Relay 2/Event
Setup/Relay 2/Event
Setup/Relay 2/Event
Setup/Relay 2/Event
Setup/Relay 2/Event
Setup/Relay 2/Event
Setup/Relay 2/Event
Setup/Relay 2/Timer
Setup/Relay 2/Timer
Setup/Relay 2/Timer
Setup/Relay 2/Timer
Setup/Relay 2/Timer
Setup/Relay 2/Timer
Setup/Relay 2/Status
Setup/Relay 3
Setup/Relay 3
Setup/Relay 3
Setup/Relay 3
Setup/Relay 3
Setup/Relay 3/Alarm
Setup/Relay 3/Alarm
Setup/Relay 3/Alarm
58
Table 11 sc100 Modbus Registers (continued)
Register
Tag Name
#
Reset
Overfeed
40113
Timer
Setpoint
40114
Phase
40116
Deadband
40117
On Max Time
40119
On Min Time
40120
Off Max Time
40121
Off Min Time
40122
Hold Type
40123
Sensor
40124
Select
Hold Mode
40125
Duration
40126
Time
Period Time
40127
Off Delay
40128
Level
40129
Source
40130
Sensor
40131
Select
Measuremen
40132
t Select
Type
40133
Transfer
40134
Setting
High Alarm
40135
Low Alarm
40137
High
40139
Deadband
Data
Length
R/W
Type
Unsigned
1
R/W
Integer
Float
2
R/W
Unsigned
1
R/W
Integer
Float
2
R/W
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Float
2
R/W
Float
2
R/W
Float
2
R/W
Description
Resets the overfeed timer
Sets the event setpoint
Sets the event phase (0 = Low; 1 =
High)
Sets the event deadband
Sets the max on time (mins)
Sets the min on time (mins)
Sets the max off time (mins)
Sets the min off time (mins)
Sets which Sensor outputs are affected
during timer on time (0 = None;
2 = Selected Sensor; 13 = All Sensors)
Selects which Sensor outputs are being
held/transferred during the timers on
time (this is used when Hold type is set
for single Sensor)
Selects hold outputs vs. set transfer
value during timers on time
Sets the timer on duration time (sec)
Sets the period between timer on events
(mins)
Sets the time the affected Sensor
outputs are held/transferred after the
timer turns off (sec)
Sets the status level which will trigger
the relay
Selects data source for this relay
(0 = None; 1 = Real Time Clock;
2 = Sensor; 4 = Calculation)
Selects Sensor source when Source =
Sensor (0 = Sensor1; 1 = Sensor2)
Selects measurement on the Sensor
(0 = Meas1 .. 3 = Meas4)
Selects the relay type (0 = Alarm; 1 =
Control; 2 = Status; 3 = Timer; 4 =
Event)
Selects the transfer value for the relays
(0 = De-energized; 1 = Energized)
Sets the high alarm setpoint
Sets the low alarm setpoint
Sets the high alarm deadband