Group Name
Tag Name
Setup/Relay 1/Timer
Hold Type
Sensor
Setup/Relay 1/Timer
Select
Setup/Relay 1/Timer
Hold Mode
Duration
Setup/Relay 1/Timer
Time
Setup/Relay 1/Timer
Period Time
Setup/Relay 1/Timer
Off Delay
Setup/Relay 1/Status
Level
Setup/Relay 2
Source
Sensor
Setup/Relay 2
Select
Measuremen
Setup/Relay 2
t Select
Setup/Relay 2
Type
Transfer
Setup/Relay 2
Setting
Setup/Relay 2/Alarm
High Alarm
Setup/Relay 2/Alarm
Low Alarm
High
Setup/Relay 2/Alarm
Deadband
Low
Setup/Relay 2/Alarm
Deadband
Setup/Relay 2/Alarm
On Delay
Setup/Relay 2/Alarm
Off Delay
Setup/Relay 2/Control
Setpoint
Setup/Relay 2/Control
Phase
Setup/Relay 2/Control
Deadband
Overfeed
Setup/Relay 2/Control
Timer
Setup/Relay 2/Control
On Delay
Setup/Relay 2/Control
Off Delay
Table 11 sc100 Modbus Registers (continued)
Register
Data
#
Type
Unsigned
40083
Integer
Unsigned
40084
Integer
Unsigned
40085
Integer
Unsigned
40086
Integer
Unsigned
40087
Integer
Unsigned
40088
Integer
Unsigned
40089
Integer
Unsigned
40090
Integer
Unsigned
40091
Integer
Unsigned
40092
Integer
Unsigned
40093
Integer
Unsigned
40094
Integer
40095
Float
40097
Float
40099
Float
40101
Float
Unsigned
40103
Integer
Unsigned
40104
Integer
40105
Float
Unsigned
40107
Integer
40108
Float
Unsigned
40110
Integer
Unsigned
40111
Integer
Unsigned
40112
Integer
Length
R/W
Sets which Sensor outputs are affected
1
R/W
during timer on time (0 = None;
2 = Selected Sensor; 13 = All Sensors)
Selects which Sensor outputs are being
held/transferred during the timers on
1
R/W
time (this is used when Hold type is set
for single Sensor)
Selects hold outputs vs. set transfer
1
R/W
value during timers on time
1
R/W
Sets the timer on duration time (sec)
Sets the period between timer on events
1
R/W
(mins)
Sets the time the affected Sensor
1
R/W
outputs are held/transferred after the
timer turns off (sec)
Sets the status level which will trigger
1
R/W
the relay
Selects data source for this relay
1
R/W
(0 = None; 1 = Real Time Clock;
2 = Sensor; 4 = Calculation)
Selects Sensor source when Source
1
R/W
= Sensor (0 = Sensor1; 1 = Sensor2)
Selects measurement on the Sensor
1
R/W
(0 = Meas1 .. 3 = Meas4)
Selects the relay type (0 = Alarm; 1 =
1
R/W
Control; 2 = Status; 3 = Timer; 4 =
Event)
Selects the transfer value for the relays
1
R/W
(0 = De-energized; 1 = Energized)
2
R/W
Sets the high alarm setpoint
2
R/W
Sets the low alarm setpoint
2
R/W
Sets the high alarm deadband
2
R/W
Sets the low alarm deadband
1
R/W
Sets the on delay time
1
R/W
Sets the off delay time
2
R/W
Sets the controller setpoint
Sets the controller phase
1
R/W
(0 = Low; 1 = High)
2
R/W
Sets the controller deadband
1
R/W
Sets the overfeed timer value (mins)
1
R/W
Sets the on delay time (sec)
1
R/W
Sets the off delay time (sec)
Modbus Protocol
Description
57