Connection Procedures - GRAPHTEC midi LOGGER GL240 Quick Start Manual

Hide thumbs Also See for midi LOGGER GL240:
Table of Contents

Advertisement

Connection Procedures

Connecting the AC Adapter
Connect the DC output of the AC adapter to the
connector indicated as "DC LINE" on the GL240.
Connect to the Analog Input Terminals
Voltage input
+
DC voltage input
-
CAUTION: Connect wire to the designated channel, where individual channels are numbered.
Connect the External Input/Output Terminals
(For logic/pulse input, alarm output, trigger input, external sampling pulse input)
* Requires B-513 pulse/logic cable.
CH 10 9 8 7 6 5 4 3 2 1
Current input
+
Shunt resister
Ex: Current signal is converted to voltage using
the shunt resister.
For 4 to 20mA current to convert the signal to 1
to 5V.
Note:Graphtec offers B-551 250 ohm precision
-
shunt resister.
< Signal assignment >
Connecting the Grounding Cable
Use a flathead screwdriver to push the button above
the GND terminal while connecting the grounding
cable to the GL240. Connect the other end of the
cable to ground.
Orange with red dotted line
Orange with black dotted line : 2
Grey with red dotted line
Grey with black dotted line
White with red dotted line
White with black dotted line
Yellow with red dotted line
Yellow with black dotted line : 4
Pink with red dotted line
Pink with black dotted line
Shielded
Thermocouple input
+
-
Compensation wire is
used if it is required.
: 1
Logic/pulse
: 3
input
: 4
: 1
: 2
Alarm output
: 3
: Trigger input/
external sampling input
GND
3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents