Error Messages - KERN DLB 160-3A Operating Instructions Manual

Dlb*a series. electronic moisture analyzer
Hide thumbs Also See for DLB 160-3A:
Table of Contents

Advertisement

Material
Tobacco
Tea, black
Pasta
Textile fibre
Theophylline
Thermoplastic PUR –
granulate
Walnut
Washing powder
Wheat spring water
Sausage casing
Toothpaste
Pulp
Cement
Sugar
Sugar beets

13 Error messages

ERR01
ERR02
ERR03
ERR05
ERR07
ERR10
ERR11
"UNLOAD":
"Err thb"
56
Weight
Drying
Sample
temperature
(g)
( °C)
1,5
100
2
105
1,5
120
0,8-1,2
85
1,5
130
15-18
80
2,8
100
2
160
2-3
90
0,2
150
2
100
2,5
130
8-12
138
4-5
138
2
130
Weight value instable or zeroing not possible.
Check the environmental conditions.
Adjustment error, e.g. instable environmental conditions
Adjustment error e.g. incorrect adjustment weight}
Data transfer not possible, as weighing value is instable. Check
the environmental conditions.
Faulty data reading
Instable display when drying starts, check environmental
conditions
Sample weight too small
Sample or sample dish wrongly positioned.
Heating top does not work, check current supply
Weighing range exceeded, placed load exceeds the capacity of
the appliance. Unload appliance.
Weighing range not reached, e.g. Dish holder / removal help is
missing.
Drying
Moisture
period
%
(approx.)
(approx.)
(min)
16
10,18
4
7,67
8
10,64
3,6
14,03
1,9
7,33
18
0,08
5,6
3,5
12
7,32
10
3,5
7,7
34,28
4,5
7,32
4-5
0,8
10
11,9
13,4
DLB_A-BA-e-1110
Solid
body
%
(approx.)
6
78,56
30,94

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents