Menu overview
Menu item
Baudrate
(siehe Kap. 9.1)
Automatic zero point
correction (see chap. 9.2)
Filter
(see chap. 9.3)
Stability display
(see chap. 9.4)
Background illumination
of the display (see chap.
9.5)
24
Display
Selection
bAud rt
br 1200
br 2400
br 4800
br 9600
Auto 0
Au0 OFF
Au0 1
Au0 2
Au0 3
Au0 3E
FiltEr
Filt 1
Filt 2
Filt 3
StAbil
Stab 1
Stab 2
Stab 3
Blt
on
off
Auto
End
Description
Auto Zero switched off
Auto Zero range ± ½ digit
Auto Zero range ± 3 digits
Auto Zero range ± 7 digits
Auto Zero range ± 7 digits in the
whole weighing range
Setting for dispensing
Sensitive and fast, very quiet
set-up location.
Robust but slow, busy set-up
location.
Standstill control fast / very quiet
set-up location
Standstill control fast + exact/
quiet set-up location
Standstill control exact /very set-
up location.
Background illumination on
Background illumination off
Backlighting automatically
switched off 3 seconds after
achieving stable weighing value.
Changes in weight or pressing
of keys will automatically result
in backlight switching on again.