9.4 Error messages
Error
message
- - - -
Maximum load exceeded
- - ol -
Err 1
Incorrect data input
Err 2
Incorrect time entry
Err 4
Zeroing range exceeded due
to switching-on balance or
pressing
(normally 4% max)
Err 5
Keyboard error
Err 6
Value outside the A/D
changer range
Err 9
Stability display does not
appear
Err 17
Taring range exceeded
Fai l h /
Adjustment error
Fai l l
Ba lo /
Battery very low
Lo ba
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
KFE-TM-BA_IA-e-1313
Description
Possible causes
Unload weighing system or reduce
preload.
Follow format "yy:mm:dd"
Follow format "hh:mm:ss"
Object on the weighing plate
Overload when zeroing
Weighing plate not installed
Damaged weighing cell
Damaged electronics
Check the environmental
conditions.
Reduce load
Repeat adjustment.
Recharge battery
29