Error Messages - KERN KFE-TM Operating And Installation Instructions

Display unit
Hide thumbs Also See for KFE-TM:
Table of Contents

Advertisement

9.4 Error messages

Error
message
- - - - ­
Maximum load exceeded
- - ol - ­
Err 1
Incorrect data input
Err 2
Incorrect time entry
Err 4
Zeroing range exceeded due
to switching-on balance or
pressing
(normally 4% max)
Err 5
Keyboard error
Err 6
Value outside the A/D
changer range
Err 9
Stability display does not
appear
Err 17
Taring range exceeded
Fai l h /
Adjustment error
Fai l l
Ba lo /
Battery very low
Lo ba
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
KFE-TM-BA_IA-e-1313
Description
Possible causes
Unload weighing system or reduce
preload.
Follow format "yy:mm:dd"
Follow format "hh:mm:ss"
Object on the weighing plate
Overload when zeroing
Weighing plate not installed
Damaged weighing cell
Damaged electronics
Check the environmental
conditions.
Reduce load
Repeat adjustment.
Recharge battery
29

Hide quick links:

Advertisement