8.2.1 Calibrate temperature

During temperature calibration only a check is carried out, i.e. no values are changed.
 Invoke service function „Tmp test", see chap. 8.2.
 Confirm by pressing the PRINT button. The current setting will be displayed. Use
the navigation buttons   to select the temperature during the check, available
35- 160°C.
 Confirm using the PRINT button, the appliance heats up to the set temperature.
After approx. 15 minutes compare the temperature displayed on the thermometer
with that of the moisture analyzer. If the two values do not match, we recommend
a temperature adjustment, see chap. 8.2.2.
 Use the ON/OFF button to finish the calibration.
 Use the ON/OFF button to return into weighing mode.
DLB_A-BA-e-1110