9.3

Filter

This menu item allows the balance to be set according to specific ambient conditions
and measuring purposes.
 In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute.
 Press MENU button repeatedly
 Acknowledge using PRINT button, the current setting is displayed.
 Use the navigation buttons   to select the desired setting.
Filter 1: Setting for dispensing
Filter 2: The balance reacts quickly and in a sensitive manner, very quiet set-up
location.
Filter 3: The balance reacts slowly and in a robust manner, busy set-up location
 Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
 Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.