Maintenance of Standard, Semi-Micro or Micro Flow Cell

When required
Tools required
Parts required
Preparations
The gaskets used in the standard and semi-micro/micro flow cell are different.
N O T E
106
If the flow cell needs repair due to leaks or contaminations (reduced light
throughput)
Two 1/4 inch wrenches for capillary connections
hexagonal key 4 mm
Tooth picks
For parts, see
"Standard Flow Cell"
page 148 or
"Micro Flow Cell"
Turn the flow off.
Remove the front cover.
Remove the flow cell, see
on page 146,
"Semi-Micro Flow Cell"
on page 150.
"Exchanging a Flow Cell"
1200 Series DAD and MWD User Manual
on
on page 103.