Optimization Overview - Agilent Technologies G1315B User Manual

Agilent 1200 series diode array and multiple wavelength detectors
Table of Contents

Advertisement

Optimization Overview

Table 11
Optimization Overview
Parameter
1 Selection of flow cell
Choose flow cell according to used column, see
2 Connection of flow cell
For flow rates from 0.5 ml/min connect column using the
zero-dead-volume fittings of the detector.
For small column i.d. (e.g 1 mm) the inlet capillary of the micro flow cell
can be connected directly to the column.
3 Setting the peak width (response time)
Use peak width according
Set the peak-width close to the width of a narrow peak of interest in
your chromatogram.
4 Setting wavelength and bandwidth
• Sample wavelength:
Never miss a peak by the use of a browser wavelength like 250 nm
with 100 nm bandwidth.
Select specific wavelength with reduced bandwidth if you need
selectivity, e.g. 250,10 nm and 360,100 nm as reference wavelength.
Set the sample wavelength to a peak or valley in the spectrum to get
best linearity for high concentrations.
• Reference wavelength:
Select the reference wavelength with broad bandwidth (30...100 nm)
wavelength range where your analytes have little or no absorbance
(e.g. sample at 254 nm, reference at 320 nm).
1200 Series DAD and MWD User Manual
Figure
Figure 25
as starting point.
Impact
• peak resolution versus sensitivity
25.
• chromatographic resolution
• peak resolution versus sensitivity versus
disk space
• sensitivity versus selectivity
• sensitivity versus linearity
• baseline drift due to RI effects.
How to optimize the Detector
5
69

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

G1365b

Table of Contents