Reflectance Standards And Gain Settings - Agilent Technologies 4200 FlexScan Operation Manual

Ftir
Hide thumbs Also See for 4200 FlexScan:
Table of Contents

Advertisement

Analyzing
Samples
The 45 degree probe is optimal for analysis of thin films greater than
1 micron in thickness. The grazing angle probe is optimal for analysis
of ultra thin films less than 1 micron in thickness. Diffuse reflectance
is specifically designed for measurement of rough surfaces that
scatter rather than reflect light. Specular reflectance is an ideal
method for the analysis of surface contamination. The grazing angle
probe should be used when very low level contamination needs to be
analyzed (nanometer thickness levels).
Each reflectance sample interface requires the use of a reference
sample for taking the background measurement prior to analyzing
the sample. The reference sample is incorporated into the 4200
FlexScan FTIR by means of a metal disk mounted into a cap that
snaps onto the tip of the reflectance probe. The 4200 FlexScan
system offers two reference caps designed for this purpose — a
diffuse gold reference cap with a rough surface finish and a stainless
steel mirror reference cap with a very smooth surface finish.
The diffuse gold reference cap should be used with the 45 degree
reflectance probe when measuring low-reflectance samples. The
mirror reference cap should be used with the 45 degree reflectance
probe when measuring thin films or highly-reflecting samples. The
mirror reference cap should always be used with the grazing angle
sample probe. The diffuse cap should always be used with the diffuse
reflectance sample interface.
Ensure that the sample makes solid contact with the entire surface area of the
CAUTION
external reflectance probe tip and is not held at an angle against the sample, and
that no movement occurs between the sample and the probe during the
measurement.

Reflectance standards and gain settings

The 4200 FlexScan FTIR is designed to measure a variety of sample
types. In order to measure the best quality spectra, it is important to
use the correct reflectance standard and to appropriately set the
detector gain. Detector gain is an electronic setting that maximizes
the detector response from the available light. The detector gain must
be set such that both the reference material and the sample are
within the linear range of the detector.
24
Agilent 4200 FlexScan FTIR Operation Manual

Advertisement

Table of Contents
loading

Table of Contents