Section 1 Safety Information - Thermo Scientific EASYpure II RF Operating Manual And Parts List

Series 1305. reservoir feed water purification system
Table of Contents

Advertisement

Section 1
Safety Information
1-2
EASYpure II RF
To avoid personal injury:
1. Do not use in the presence of flammable or combustible materials;
fire or explosion may result. This device contains components which
may ignite such materials.
2. Do not use in the presence of highly corrosive substances such as
bleach or acid baths; fire may result.
3. This device is to be used with water feeds only. Sanitizing/cleaning
agents must be used in compliance with instructions in this manual.
Failure to comply with the above could result in explosion and
personal injury.
4. Avoid splashing disinfecting solutions on clothing or skin.
5. Ensure all piping connections are tight to avoid chemical leakage.
6. Carefully follow manufacturer's safety instructions on labels of
chemical containers and material safety data sheets.
7. Refer servicing to qualified personnel.
8. Ensure adequate ventilation.
9. Depressurize system prior to opening cartridge access door or
removing top cover.