Section 7 Initial Operation - Thermo Scientific EASYpure II RF Operating Manual And Parts List

Series 1305. reservoir feed water purification system
Table of Contents

Advertisement

Section 7
Initial Operation
Filling and Rinse Up
(continued)
7-2
EASYpure II RF
8. Place suitable container under draw-off valve. Open draw-off valve.
9. Press START and rinse one complete reservoir volume of water
through the cartridges to drain.
10. Close draw-off valve.
11. Remove hose barb connection from draw-off valve. Keep hose barb for
future use.
12. Remove the new 0.2 micron final filter assembly from its bag and
insert it into the draw-off valve. Gently tighten, turning the filter to
the right.
Note It is suggested that Teflon® tape be applied to the threads of the 0.2
micron final filter to ensure a tight seal.
13. Refill the reservoir with pretreated water.
14. Open the draw-off valve and flush second reservoir volume of water
through the 0.2 micron final filter.
Note For more demanding applications where low TOC water is required,
a third reservoir volume rinse of cartridges and filter may be necessary.