Normal Operation; Operation; Water Draw-Off - Thermo Scientific EASYpure II RF Operating Manual And Parts List

Series 1305. reservoir feed water purification system
Table of Contents

Advertisement

Operation

Water Draw-Off

Thermo Scientific

Normal Operation

Section 8
Note When the unit is first put in the Run Mode, the display will show
10.0 for a few seconds. This is an arbitrary number that indicates the unit
is running. Any number that appears after 10.0 indicates purity.
1. Remove the Ventgard cap. Do not allow water to enter the Ventgard
cap.
2. Carefully pour pretreated feedwater into the feedwater reservoir until
the water is level with the top of the sight gauge. Replace the Ventgard
cap.
3. Turn main power on at power entry module.
4. Press the "START" button on the front of the EASYpure II RF. The
EASYpure II RF's pump will begin to run and the purity meter will
display "10.0" followed by the number indicating the resistivity of the
water in megohm-cm.
5. Allow the water's resistivity to rise to the desired purity before drawing
off water.
1. Remove the protective cap from the 0.2 micron final filter bell.
2. Depress the draw-off lever.
3. When draw off is complete, lift the draw-off lever and replace the
protective cap on the 0.2 micron final filter bell.
s
EASYpure II RF
8-1