Unit Shutdown; Fuse Replacement; Ventgard Cap Replacement - Thermo Scientific EASYpure II RF Operating Manual And Parts List

Series 1305. reservoir feed water purification system
Table of Contents

Advertisement

Section 9
Maintenance and Servicing
Ventgard Cap
Replacement

Fuse Replacement

Unit Shutdown

9-8
EASYpure II RF
The purifying media and filter in the Ventgard filter element have a limited
capacity. Therefore, the Ventgard cap should be replaced every 120 days.
The Ventgard cap is shipped as a complete unit; replacement involves
simply removing the new Ventgard cap from its plastic storage bag and
placing it on the reservoir. A Ventgard cap can be stored in a cool, dry
place almost indefinitely, provided its plastic storage bag has not been
opened.
1. Disconnect the EASYpure II RF from the power supply.
2. Remove the power cord from the power entry module.
3. Pull out the fuse drawer located in the power entry module on the back
right side of the unit.
4. Remove old fuses and replace with fuses of the same type and rating.
See Replacement Parts list.
5. Replace fuse drawer.
6. Replace power cord.
7. Reconnect unit to power supply
Warning Replace fuses with those of the same type and rating.
If the EASYpure II RF will be inactive for a period up to a month, place
the unit in Standby. For periods of time greater than a month, disconnect
water and power sources, drain unit, close customer supplied water valve,
remove and discard cartridges and final filter. See Sanitization, Cartridge
Replacement and 0.2 Micron Final Filter Replacement procedures for
reactivation.