Badger Meter B2800 User Manual

Flow monitor panel mount standard
Hide thumbs Also See for B2800:

Advertisement

Quick Links

Flow Monitor
B2800 Panel Mount Standard
User Manual
DSY-PM-00278-EN-04 (November 2017)

Advertisement

Table of Contents
loading

Summary of Contents for Badger Meter B2800

  • Page 1 Flow Monitor B2800 Panel Mount Standard User Manual DSY-PM-00278-EN-04 (November 2017)
  • Page 2: Table Of Contents

    Flow Monitor, B2800 Panel Mount Standard CONTENTS Scope of This Manual Unpacking and Inspection Safety Terminology and Symbols Considerations Introduction Installation Operating The Monitor Buttons Modes Programming Programming Mode Run Mode Additional Scaling Parameters Flow 4 mA Setting Flow 20 mA Setting 4…20 mA Calibration...
  • Page 3: Scope Of This Manual

    Considerations The installation of the B2800 monitor must comply with all applicable federal, state, and local rules, regulations, and codes EXPLOSION HAZARD - SUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY FOR CLASS I, DIVISION 2.
  • Page 4: Introduction

    INTOODUCTION The B2800 flow monitor is a state-of-the-art digital signal processing flow monitor, designed to provide exceptional flexibility at an affordable price Though it is designed for use with Blancett flow meters, this display can be used with almost any flow...
  • Page 5: Operating The Monitor

    Programming Choices Programming Keys Units Display and Programming Choices Figure 3: B2800 panel mount flow monitor Buttons MENU Switches to PROGRAM mode Scrolls forward through the parameter options and increments numeric variables Scrolls backward through the parameter options and moves...
  • Page 6: Programming

    Programming POOGOAMMING Each turbine flow meter is shipped with either a K-factor value or frequency data If frequency data is provided, the data must be converted to a K-factor before programming the monitor K-factor information, when supplied, can usually be found on the neck of the flow meter or stamped on the flow meter body The K-factor represents the number of pulses per unit of volume The K-factor is needed to program the monitor readout Programming Mode...
  • Page 7: Run Mode

    Programming Select the Units of Measure The monitor is programmed with five common rate/total unit options The monitor shows the rate/total unit that the display is currently set for If the current selection is correct, press ENTEO once to advance to the DISPLAY FUNCTION parameter To change the unit, press UP or OIGHT to scroll to the correct rate unit Press ENTEO to save and advance to the DISPLAY FUNCTION parameter Selection...
  • Page 8: Additional Scaling Parameters

    Additional Scaling Parameters ADDITIONAL SCALING PAOAMETEOS OTEE: The programming instructions below are only available for loop powered units Battery powered units do not include these programming parameters Flow 4 mA Setting Zero is the default flow rate at the 4 mA setting If the current selection is correct, press ENTEO to advance to the Flow 20 mA parameter To change the setting, press UP to increment the flashing digit and press OIGHT to move to the next digit Press ENTEO to save and advance to the Flow 20 mA parameter Flow 20 mA Setting...
  • Page 9: Battery Replacement

    ADDITIONAL INPUT OPTIONS The B2800 flow monitor is capable of receiving magnetic pickup input (small signal sine wave) or a contact closure input (pulse) Since most Blancett turbine flow meters utilize a magnetic pickup, the B2800 flow monitor is shipped configured for...
  • Page 10: Specifications

    Specifications SPECIFICATIONS Power Supply Options Battery Powered One D size, 1 5V alkaline battery Loop Powered Optional 4…20 mA loop power Alphanumeric Oate and Total Display Eight digit, 0 50 inch high numeric display Eight character, 0 25 inch high alphanumeric display Fixed or toggle modes of operation for flow rate and totalizer display Accuracy ±0 1%...
  • Page 11: Dimensions

    Dimensions DIMENSIONS 4 25 in (107 9 mm) 4 25 in (107 9 mm) 0 75 in (19 0 mm) 2 50 in (63 5 mm) 4 54 in (115 3 mm) TOOUBLESHOOTING GUIDE Issue Oemedy No LCD display • Check battery voltage Should be 1 5V DC Replace if low or bad •...
  • Page 12: Default K-Factor Values

    The Americas | Badger Meter | 4545 West Brown Deer Rd | PO Box 245036 | Milwaukee, WI 53224-9536 | 800-876-3837 | 414-355-0400 México | Badger Meter de las Americas, S.A. de C.V. | Pedro Luis Ogazón N°32 | Esq. Angelina N°24 | Colonia Guadalupe Inn | CP 01050 | México, DF | México | +52-55-5662-0882 Europe, Eastern Europe Branch Office (for Poland, Latvia, Lithuania, Estonia, Ukraine, Belarus) | Badger Meter Europe | ul.