    Flow Monitor B2800 Standard User Manual DSY-PM-00276-EN-04 (March 2017)
  • Page 2: Table Of Contents

    Flow Monitor, B2800 Standard CONTENTS Scope of This Manual � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � 3 Unpacking and Inspection �...
  • Page 3: Scope Of This Manual

    Scope of This Manual SCOPE OF THIS MANUAL This manual is intended to help you get the B2800 flow monitor up and running quickly� MPOOTANT Read this manual carefully before attempting any installation or operation. Keep the manual accessible for future reference. UNPACKING AND INSPECTION Upon opening the shipping container, visually inspect the product and applicable accessories for any physical damage such as scratches, loose or broken parts, or any other...
  • Page 4: Introduction

    Introduction INTOODUCTION The B2800 flow monitor is a state-of-the-art digital signal processing flow monitor, designed to provide exceptional flexibility at a very affordable price� Though designed for use with Blancett flow meters, this display can be used with almost any flow meter producing a low amplitude AC output or contact closure signals�...
  • Page 5: Operating The Monitor

    Operating the Monitor OPEOATING THE MONITOO Rate / Total Display Mode Indicator Programming Choices PROGRAM RELAY1 RELAY2 Units Display Programming Keys Programming Choices Figure 1: B2800 flow monitor Buttons MENU Switches to PROGRAM mode Scrolls forward through the parameter options and increments numeric variables Scrolls backward through the parameter options and moves the active digit OIGHT...
  • Page 6: Programming

    Programming POOGOAMMING Each turbine flow meter is shipped with either a K-factor value or frequency data� If frequency data is provided, the data must be converted to a K-factor before programming the monitor� K-factor information, when supplied, can usually be found on the neck of the flow meter or stamped on the flow meter body�...
  • Page 7 Programming Example If the meter was calibrated in US gallons, use PUL/GAL� Press ENTEO to save the K-factor unit and advance to the K FACTOR parameter� OTEE: Unless otherwise specified, Blancett turbine flow meters are supplied with K-factors measured in pulses per gallon (PUL/GAL) which will automatically convert to your desired units of measure�...
  • Page 8: Run Mode

    Programming Totalizer Pulse Output The pulse output parameter can be enabled or disabled� When enabled this output generates 20 mS duration pulse for every time the least significant digit of the totalizer increments� The amplitude of the pulse is dependent on the voltage level of the supply connected to the pulse output and is limited to a maximum 30V DC�...
  • Page 9: Additional Scaling Parameters

    Additional Scaling Parameters ADDITIONAL SCALING PAOAMETEOS OTEE: The programming instructions below are only available for loop powered units� Battery powered units do not include these programming parameters� Flow 4 mA Setting Zero is the default flow rate at the 4 mA setting� If the current selection is correct, press ENTEO to advance to the Flow 20 mA parameter�...
  • Page 10: Ma Test

    Additional Scaling Parameters 4…20 mA Test The monitor contains a diagnostic routine that allows the simulation of mA values between 4…20 to check output tracking� At the 4-20TEST prompt press UP to increment the simulated mA output or OIGHT to decrement the value in increments of 1 mA�...
  • Page 11: Additional Input Options

    Additional Input Options ADDITIONAL INPUT OPTIONS The B2800 flow monitor is capable of receiving magnetic pickup input (small signal sine wave) or a contact closure input (pulse)� Since most Blancett turbine flow meters use a magnetic pickup, the B2800 flow monitor is shipped configured for magnetic pickup input� To change to a contact closure input, remove JP2 from the top two pins and jumper them to the bottom two pins�...
  • Page 12: Default K-Factor Values

    Default K-factor Values DEFAULT K-FACTOO VALUES Meter Size Default K-factor Lower Limit Upper Limit 3⁄8 in� (9�53 mm) 20,000 16,000 24,000 1/2 in� (12�7 mm) 13,000 10,400 15,600 3/4 in� (19�05 mm) 2750 2200 3300 7/8 in� (22�23 mm) 2686 2148 3223 1 in�...
  • Page 13: Installation Drawing

    Installation Drawing INSTALLATION DOAWING Battery Powered March 2017 DSY-PM-00276-EN-04 Page 13...
  • Page 14: Loop Powered

    Installation Drawing Loop Powered Page 14 DSY-PM-00276-EN-04 March 2017...
  • Page 15: Troubleshooting Guide

    Troubleshooting Guide TOOUBLESHOOTING GUIDE Issue Oemedy • Check battery voltage� Should be 1�5V DC� Replace if low or bad� No LCD display • Check connection from meter pickup to display input terminals� No rate or total • Check turbine meter rotor for debris� Rotor should spin freely� displayed •...
  • Page 16: Specifications

    Specifications SPECIFICATIONS Power Supply Options Battery Powered One D size, 1�5V alkaline battery Loop Powered Optional 4…20 mA loop power Alphanumeric Oate and Total Display Eight digit, 0�75 inch high numeric display Eight character, 0�38 inch high alphanumeric display Fixed or toggle modes of operation for flow rate and totalizer display Accuracy ±0�1% Temperature Drift...
  • Page 17: Replacement Parts

    Replacement Parts OEPLACEMENT PAOTS Component Part Number Keypad B260713 Battery B280601 Battery Tie Wrap B228036 Pickup Cable B222-121 Desiccant Bag B260630 PVC Union B220016 PVC Reducer Bushing B220056 Rubber Washer B228207 Steel Lock Washer B220018 PCB Shield (battery units) B280603 Desiccant Shield B280680 Cord Grip...
  • Page 18: Mounting Options And Dimensions

    Mounting Options and Dimensions MOUNTING OPTIONS AND DIMENSIONS Meter Mount G dia 9�25 in� 7�00 in� 5�75 in� 4�00 in� 3�45 in� 1�50 in� 0�875 in� (235�0 mm) (177�8 mm) (146�0 mm) (101�6 mm) (87�6 mm) (38�1 mm) (22�2 mm) Oemote Mount .25 in.
  • Page 19: Swivel Mount

    Mounting Options and Dimensions Swivel Mount G dia 12�25 in� 7�00 in� 5�75 in� 4�00 in� 3�45 in� 1�50 in� 0�875 in� (311�2 mm) (177�8 mm) (146�0 mm) (101�6 mm) (87�6 mm) (38�1 mm) (22�2 mm) March 2017 DSY-PM-00276-EN-04 Page 19...
  • Page 20 The Americas | Badger Meter | 4545 West Brown Deer Rd | PO Box 245036 | Milwaukee, WI 53224-9536 | 800-876-3837 | 414-355-0400 México | Badger Meter de las Americas, S.A. de C.V. | Pedro Luis Ogazón N°32 | Esq. Angelina N°24 | Colonia Guadalupe Inn | CP 01050 | México, DF | México | +52-55-5662-0882 Europe, Eastern Europe Branch Office (for Poland, Latvia, Lithuania, Estonia, Ukraine, Belarus) | Badger Meter Europe | ul.