AI 3000/AS 3000 II Menu

Thermo Scientific
To set the parameters of the AI 3000/AS 3000 II method, In Main Menu move the cursor
until Autosampler is displayed, then press Enter to open the menu. See
Table 11. Control Through the FOCUS GC: Autosampler Menu (Sheet 1 of 2)
Menu
AUTOSAMPLER
Sample volume
Sample rinses
Plunger strokes
Pre wash solvent
Pre wash cycles
Post wash solvent
Post wash cycles
Viscous sample
Range
Comment
This line displays the menu title.
0–5 μL, 0-2.5 μL
This parameter allows to set the sample
with increments of
quantity to be injected.
0.1 μL
0–15
This parameter allows to set the number of
syringe pre-washings with the sample.
0–15
This parameter allows to set the number of
plunger strokes to eliminate air bubbles
forming during the sample withdrawal.
A, B, C, D,
This parameter allows to select the vial, or
A+B,
combination of two vials, containing the
C+D
washing solvent.
0–15
This parameter allows to set the number of
syringe pre-washings with the selected
solvent.
A, B, C, D
This parameter allows to select the vial, or
A+B
combination of two vials, with the washing
C+D
solvent.
0–15
This parameter specifies the number of
syringe post-washings with the solvent
selected.
Yes, No
This parameter sets the sample drawing
speed from the vial as a function of the
sample viscosity.
AI 3000/AS 3000 Sampling System - Operating Manual
4
AI 3000/AS 3000 II Control
Control Through the FOCUS GC
Table 11
• Select No (default value) if the sample
has low viscosity.
• Select Yes when the sample has high
viscosity.
for details.