Tests and Adjustments
Performance Tests
Performance Tests
The performance tests verify the electrical performance of your PNA. These
performance tests are included in the analyzer's firmware with Option 897 and
Option 898. Your analyzer is automatically configured for each individual test.
The model numbers of the equipment used by these performance tests are
specified under
There are nine tests in the Option 897 or 898 performance test package:
—
—
—
—
—
—
—
—
—
Source Power Accuracy Test
Function of the Test: To confirm the accuracy of the source output power of
your network analyzer over its full frequency range.
Specification Tested: Test Port Output–Power Level Accuracy
Equipment Used:
— Power meter
— Power sensors
— Any necessary adapters
Description of the Test:
1. The analyzer is Preset.
2. The analyzer is set up for a CW reflection measurement on the test port to
3. A power sensor is connected to the test port, using any necessary
4. The analyzer frequency is set to the desired value.
5. The power meter correction table is set to the same frequency.
Keysight N5224A/25A Service Guide
"Required Service Test Equipment" on page
Source Power Accuracy Test
Source Maximum Power Output Test
Source Power Linearity Test
Frequency Accuracy Test
Trace Noise Test
Receiver Compression Test
Noise Floor Test
Calibration Coefficients Test
Dynamic Accuracy Test
be measured.
adapters.
2-4.
3-31