Theory of Operation

Receiver Group Operation

Receiver Group Operation
The receiver group measures and processes the input signals into digital
information for processing and eventual display.
Figure 5-7 on page 5-26
functional group for 2-port and 4-port models.
In this section the following assemblies are described:
A46–A49 35-dB Receiver Step Attenuators
A step attenuator is placed in the signal path of each of the A, B, C, and D
receiver inputs for 4-port models and A and B receiver inputs for 2-port
models.
These 35-dB step attenuators provide power control for the input signals to the
mixer bricks. They are electro-mechanical step attenuators that provide 0 to 35
dB of attenuation in 5-dB steps. These attenuators are controlled by the A21
CPU board.

A37 Reference Mixer Switch

The A37 reference mixer switch is placed in the R1 reference signal path
allowing this reference signal to be switched in and out of the signal path when
an external mixer is being used in test configuration.
An external mixer is placed in measurement configuration between
REFERENCE 1 SOURCE OUT and RCVR R1 IN where there would normally be a
front panel jumper. The A37 reference mixer switch can then be used to switch
this external mixer in and out of the measurement configuration without having
to manually connect/disconnect the external mixer and remove/replace the
front panel jumper.

A27 and A28 Mixer Bricks

Each of these assemblies contain four identical amplifiers, mixers, and filters
for a total of eight of each. For 2-port models, only the A27 mixer brick is
present; the A28 mixer brick is omitted since only four receivers are needed.
For frequencies at or above 53 MHz, the test signals (receivers A, B, C, and D
for 4-port models and A and B for 2-port models) and the reference signals
(receivers R1, R2, R3, and R4 for 4-port models and R1 and R2 for 2-port
5-22
A46–A49 35-dB Receiver Step Attenuators
A37 Reference Mixer Switch
A27 and A28 Mixer Bricks
A24 IF Multiplexer Board
A16 SPAM Board (Analog Description)
are simplified block diagrams of the receiver
Keysight N5224A/25A Service Guide
Figure 5-6 on page 5-25
and

Advertisement

Table of Contents
loading

This manual is also suitable for:

N5225a

Table of Contents