Tests and Adjustments
System Verification
4. If the analyzer still fails the test, check the measurement calibration by
viewing the error terms as described in
page
5. Refer to
Figure 3-11
System Verification Failure Flowchart
Interpreting the Verification Results
Figure 3-12
Data selected in the Printer Output area of the System Verification dialog box.
At the top of the printed output is the name of the device, the serial number of
the device, and the date tested.
Keysight N5224A/25A Service Guide
8-6.
Figure 3-11
for additional troubleshooting steps.
shows an example of typical verification results with Print Tabular
"Accessing Error Terms" on