Non-Standards Compliant Verification - Keysight Technologies N5224A Service Manual

2-port and 4-port pna microwave network analyzers, 10 mhz to 43.5 / 50 ghz
Hide thumbs Also See for N5224A:
Table of Contents

Advertisement

Tests and Adjustments

Non-Standards Compliant Verification

Non-Standards Compliant Verification
To meet the criteria for non-standards compliant verification, perform the
preliminary checks and the performance tests while stopping to troubleshoot.
Refer to
encountered without continuing to other tests. After you troubleshoot, repair,
and adjust, repeat the last failed portion and generate a new set of data.
Figure 3-2
Non–Standards Compliant Verification Flowchart
Keysight N5224A/25A Service Guide
Figure 3-2
for test flow. Troubleshoot and repair the first problem
3-11