Keysight Technologies N5224A Service Manual page 102

2-port and 4-port pna microwave network analyzers, 10 mhz to 43.5 / 50 ghz
Hide thumbs Also See for N5224A:
Table of Contents

Advertisement

Troubleshooting
Measurement System Troubleshooting
— Connect an Open or Short standard from a mechanical calibration kit to
each test port (use adapters if necessary).
— Press UTILITY
Receiver Display
— For 2-port analyzer models, traces A, B, R1, and R2 are displayed in four
separate data windows as shown in
the traces in these windows can help you to isolate the faulty assembly.
Figure 4-7
Typical 4-Receiver Display for 2-Port Models
— For 4-port analyzer models, traces A, B, C, D, and R1, R2, R3, and R4 are
displayed in eight separate data windows as shown in
Identifying discrepancies of the traces in these windows can help you to
isolate the faulty assembly.
4-30
Service
System
, then
.
Figure
Keysight N5224A/25A Service Guide
Utilities
, then
, then
4-7. Identifying discrepancies of
Figure
4-8.

Advertisement

Table of Contents
loading

This manual is also suitable for:

N5225a

Table of Contents