ii.
Click OK to save the name. Verify that the location name is created and displayed.
iii.
Select the location (ensure it is highlighted).
b.
If the desired location has been created and is displayed, select the location (ensure it is highlighted).
3.
Click the Comm. Port drop-down list.
4.
Select the laptop communication port used.
5.
Leave Baud Rate at 9600 (factory default). It is assumed that the unit is new and is still configured with its factory defaults.
6.
Click Setup LevelMaster. The New LevelMaster Setup window displays (Figure 40).
7.
Select I have only one LevelMaster connected to communication port and I want to set it up.
8.
Click Setup LevelMaster.
9.
When prompted to add the LevelMaster to the list, click Yes.
10. When prompted for the LevelMaster name, type up to 15 characters and click OK. The name will be stored in the location
file and associated with the newly created location.
11. Observe the messages displayed on the monitor screen (Figure 41). Verify that the communication is successful. The
screen should indicate the number of tries and 100% success. If errors are displayed and communication is not
established, verify that the cables and adapters are properly attached and that the correct laptop port and baud rate were
selected. If unable to communicate, see the contact information on the back of this manual for support.

Figure 40: New LevelMaster Setup

LevelMaster 7100 User manual 2105835 Rev AA | 43