Figure 52: Adding A Variable (Measurement) To Trend - ABB LevelMaster 7100 User Manual

Liquid level sensor with batteryless float support
Hide thumbs Also See for LevelMaster 7100:
Table of Contents

Advertisement

8.
Repeat steps 6 and 7 for additional LevelMaster data items or variables.
9.
Click Send to save the configuration.
10. Test the configuration in the next procedure.
Test configuration by collecting the data
To verify that the variables were defined correctly:
1.
Select the collect icon on the top tool bar.
2.
At the collection screen, on the navigation tree, expand TOTALFLOW.
3.
Expand Trend files and select the LevelMaster file.
4.
Select Screen as ouput and click Collect.
5.
Select the tab for the preferred view (graph or table) (Figure 53).Verify that the defined variables and their values are
displayed as expected. Each variable defined in the trend file can be edited and corrected, or the variable can be deleted
and redefined with the correct information from the Trend System screen.
LevelMaster 7100 User manual 2105835 Rev AA | 58

Figure 52: Adding a variable (measurement) to trend