4 Sensor setup – first-time installation
IMPORTANT (NOTE)
Perform this procedure when a new / replacement
sensor is connected to the transmitter for the first time
only. For existing sensors, see Section 5, page 8.
To perform a first-time installation (Easy Setup menu):
1.
Connect a new or replacement sensor to the transmitter's
EZLink connector – see transmitter Operating instructions
OI/AWT440-EN.
The following prompt is displayed identifying the new /
replacement sensor (S1 to S4):
ATS400
New Sensor Sx
Detected
Start Easy
Setup ?
2.
To enter Easy Setup level, press the
(below the
icon).
The Easy Setup start screen is displayed:
Menu
Easy Setup
Exit
Press the
key (below the Select prompt).
3.
To enter Easy Setup level, press the
(below the Select icon).
4.
Press the
key (below the Edit prompt) to change the
default value to the required value / selection.
5.
Press the
key (below the Next prompt) to accept the
value / selection displayed and advance to the next
configuration parameter.
00-00-2014
key
Select
key
The following Configuration parameters are set at Easy
Setup level:
— Tag
— PV Type
— Units
— Range High
— Range Low
— Clean Interval
— Filter Type
— Analogue outputs and alarms
IMPORTANT (NOTE)
Refer to Section 5.1, page 9, for parameter details – not
all parameters in Section 5.1 are displayed at Easy Setup
level.
6.
Continue with configuration of the required parameters.
On completion the Easy Setup start screen is displayed:
Menu
Easy Setup
Exit
7.
To exit Easy Setup, press the
prompt) to display the Operator Page.
Pressing the
key (below the Select prompt) re-enters the
Easy Setup level where parameters can be reviewed or
modified after 1st time connection.
After completing the Easy Setup level, pressing the
or
key enters the Advanced Configuration level, where
all available sensor and transmitter parameters can be
reviewed or modified.
IMPORTANT (NOTE)
To re-configure an existing sensor (after first-time
installation), enter the Configuration level (see
Section 5.1, page 9) via the Operator Page – refer to
transmitter Operating instructions
Operator Page details and navigation.