6
3

Sensor setup menu

Dissolved oxygen
Menu
Comment
Tag
Enter an alphanumeric sensor tag (16 characters maximum) to identify the sensor
on the Operator Pages.
Measurement Type
Select the required probe type:
• Dissolved Oxygen / % Saturation
Note. If a change is made the I/O sources are reset.
Units
Select the measurement units:
• mg/L / ppm
PV Resolution
Select the PV Resolution:
• Normal / High
Range High
Set the span value in Chart and Bargraph views.
Range Low
Set the zero value in Chart and Bargraph views.
Filter Type
Select the signal filtering type:
• None / Low / Medium / High
Salinity Unit
Select the required salinity units:
• PSU (Practical Salinity Units) or ppt (parts-per-thousand).
Salinity Correction
Required when monitoring water containing high quantities of dissolved salts:
• enter the required value between 0 and 42 Practical Salinity Units (PSU).
• leave at the default value of 0 PSU if salinity correction is not required.
Pressure Unit
Select the required barometric pressure units:
• mbar / mmHg
Barometric Pressure
Barometric pressure compensation.
Set the local barometric pressure to 506 to 1114 mbar (380 to 835 mmHg).
If the barometric pressure is unknown, leave at the default sea-level value of
1013 mbar (760 mmHg).
Slope Limit
0.85 to 1.20
Offset Limit
-0.2 to 0.2
Reset To Defaults
Select to reset all Sensor Setup parameters to their default values.
A D S 4 2 0 | D I S S O LV E D OX YG E N S E N S O R | O I/A D S 4 2 0 - E N R E V. B
Default
TAG1
Dissolved Oxygen
ppm
Normal
50 ppm (200 %)
0
None
PSU
0 PSU
mbar
1013 mbar
N/A
N/A
N/A