Step 2: Install The Stage Unit - Thermo Scientific Versette Brief User Manual

Table of Contents

Advertisement

Step 2: Install the stage unit

1.
Manually move the stage locking pin to the rear of the base platform.
2.
Carefully place the stage assembly onto the base platform and align to the four corner pins as
shown. Use a 4 mm hex wrench to secure the stage assembly onto the base platform with
M6 x 12 button head screws.
3.
Attach the safety shield cable and the stage control cable to the system connections, as shown.
The control cable is "keyed" to match the pins in the connector, and must be pushed into place.
4.
Engage the stage into the stage locking pin:
With one hand, reach over the stages and push the locking pin forwards towards yourself as you
gently push the stage assembly towards the rear of the system.
Be sure to slide the locking pin forward and hold firmly while pushing the stages to the rear. The
stage assembly should now engage the base pin and belt and slide with firm pressure, in and out.
Do not slam or rattle the stage assembly. Simply press firmly.
Stage power/control connection
Safety shield cable
7