Operation

Step 1: Perform an operational check

Complete the following procedure to confirm the correct functioning of the system prior to normal use.
1.
Verify that the safety shields are in place and closed.
2.
Verify that the power cable is connected to an appropriate power supply.
3.
Switch the instrument on using the power switch. The stages will move to the home positions and
the dispense mechanism lifts to the up position. Wait for the touchscreen to initialize and display
the main menu.

Step 2: Install ControlMate Software

Computer minimum requirements:

Computer running Microsoft
Screen resolution set to at least 1024 x 768
CD-ROM, removable drive, or network drive for access to installation software

ControlMate Software installation process

ControlMate Software can be used from a CD, a flash drive, or installed from a common directory or
server.
1.
Locate and double-click on the setup.exe file to launch the installation program.
2.
At the InstallShield
3.
Read the License Agreement, select "I accept the terms in the license agreement" to agree and
continue to install the software, then click "Next".
4.
Wait for the ControlMate files to copy to your installation directory, then select "Finish" when
displayed, to exit the wizard.
5.
Refer to the ControlMate User Manual (Cat. No. 110762) for user instructions.

Step 3: Connect the Versette to a computer running ControlMate Software

Refer to the Versette ControlMate 1.2.0 User Manual for details on any of the following steps.
®
1.
Use Windows
Control Panel to set the computer to work with ControlMate:
- Set Region to "English" (ControlMate uses English conventions for all numerical entries).
- Disable hibernation mode.
- Disable sleep mode.
2.
Connect the computer to the Versette with an RS-232 cable:
Do not use an RS-232 cable longer than 3 m when connecting the computer to the system.
Serial connector details: RS-232C, 115,200 bps, 8 data bits, 1 stop bit, and parity: none