2.0 Alarm settings 'High'/'Low'
To set High or Low pressure warning limits proceed as
follows:
•
Press the
key to switch on the PC6-IDOS.
ON/OFF
•
Select the 'Set' menu by pressing the
•
Select 'High' or Low' by pressing the
press the
key.
ENTER
ZERO
•
Select the digits to be set by moving the cursor with the
keys and pressing the
•
Set the number to the required value using the
keys and then press the
•
Repeat the previous two steps to change other digits for
the alarm value.
•
Once finished setting the alarm limit value press the
key to return to the pressure display.
A bleeping warning ( 2 Hz ) and the 'Outside Limits'
message will signal that the measured pressure is lower
than the user set low alarm value or greater than the user
set high alarm value. The bleeping ceases when the
pressure is read between the two values.
Note:
The PC6-IDOS will not allow pressure values to be
entered greater than 5% above the maximum working range.
K0354, Iss. 1
key twice.
MENU
key and then
key.
ENTER
ZERO
key.
ENTER
ZERO
MENU
Page
21