GE Druck PC6-IDOS User Manual page 35

Pressure calibrator (bench/panel mount version)
Table of Contents

Advertisement

2.6 Log
Switch on PC6-IDOS.
Press the
key twice, scroll to and s
MENU
press
.
ENTER
ZERO
The display gives the option to log a file, view a file or
change the log status as shown below:
2.6.1 Log:
Select 'Log' using the
key.
Presented with the sub-menu, enter a filename and
record count if no files are present, or select a file to log
into.
filename
To enter a filename:
Using the
and press the
ENTER
ZERO
Select the characters/numbers to be used to name the
file by scrolling up or down at each position and press
the
key when completed.
ENTER
ZERO
Press the
key to move on to the record count.
MENU
K0354, Iss. 1
elect 'Log' and
key and then press the
'N' = Normal Log
'C' = Calibration procedure log (requires 'SiCal PRO'
software) *
record count
(max. 8 characters)
keys, move to the filename position
key.
As with the filename, define the record count and then
press the
MENU
If a file has already been set up with the same name
then the 'Overwrite' option screen will be displayed.
Start logging by pressing the
on page 26).
The
symbol will appear to show file is open.
ENTER
ZERO
2.6.2 View:
Select 'View' from the log menu using the
then press the
Select the file to view using the
then press the
* Please see Appendix 2, page 58 for details of Calibration
Procedures.
key when completed.
PRINT
key. (see 2.5 'Event'
LOG
key.
ENTER
ZERO
key.
ENTER
ZERO
key and
keys and