Model
PC6-B-0.07-C
G
PC6-B-0.2-C
G
PC6-B-0.35-C
G
PC6-B-0.7-C
G
PC6-B-1-C
G
PC6-B
-2-C
G or A
PC6-B-3.5-C
G
PC6-B-7-C
G or A
PC6-B
-10-C
G
PC6-B
-20-C
G or A
PC6-B-35-C
G
PC6-B-70-C
G
PC6-B-100-C
G
PC6-B-200-C
G
PC6-B-350-C
SG
PC6-B-700-C
SG
Other Pressure ranges are available to special order.
SG = Sealed Gauge,G = Vented Gauge, A = Absolute
† 0.075% FS accuracy, 0 to 50 C
Measurement
Range
Voltage *
0 to 100mV
(Auto Range)
0 to 5V
0 to 50V
Current *
0 to 50mA
Auto Range
0 to 25mA
Sourcing
Range
Voltage *
0 to 10V
Current *
0 to 50mA
* Temperature coefficient ± 0.005% of reading / °C w.r.t. 20°C (68°F)
K0354, Iss. 1
SPECIFICATIONS
Pressure Range
Bar
psi
-0.07 to 0.07
-1.0
to
1.0
-0.2 to 0.2
-3.0 to 3.0
-0.35 to 0.35
-5.0 to 5.0
-0.7 to 0.7
-10 to 10
-1 to 1
-15 to
15
-1 to 2
-15 to
30
-1 to 3.5
-15 to
50
-1 to 7
-15 to
100
-1 to 10
-15 to
150
-1 to 20
-15 to
300
-1 to 35
-15 to
500
0 to 70
0 to
1,
000
0 to 100
0 to
1,500
0 to 200
0 to
3,
000
0 to 350
0 to
5,
000
0 to 700
0 to 10,000
° (32 to 122 °F)
Input resistance
Resolution
Accuracy
0.025% FS
1 V
µ
0.025% FS
0.1mV
0.025% FS
1mV
0.025% FS
1 A
µ
0.025% FS
Resolution
Accuracy
0.2mV
0.025% FS
0.025% FS
1 A
µ
Overload Pressure
Accuracy (FS)
Resolution
up to
EPM
0.05%
0.05%
Pressure Units
0.05%
0.05%
Overload Warning
0.025%
Temperature Display
0.025%
0.025%
Pressure Switch Test
0.025%
0.025%
Operating Temperature
0.025%
Calibration Temperature
0.025%
Calibrated Temp. range
0.025%
Storage Temperature
0.025%
Temperature Effects
0.025%
0.025%
RS232 Parameters
0.025%
Zero Reset
Data Capacity
Display
Humidity
1 MΩ
Resolution Select
1 M
Power Supply
1 M
6
Voltage Output
Max. Load
20mA
330 Ohm
15% FS
6 Digits
External Pressure Modules. See Appendix 1 for
EPM details.
bar, mbar, MPa, kPa, psi, kg/cm², atm, inH 0,
mH 0, mmH 0, inHg, mmHg, UNIT1, UNIT2 (user
2
2
defined)
Flashing display and audible tone at 115% of FS
Temperature of pressure sensor
(Accuracy: ±1 °C (1.8 °F) )
Status OPEN/CLOSED
0 to 50 °C (32 to 122 °F)
20 ±2 °C (68 ±3.6 °F)
5 to 45 °C (41 to 113 °F)
-20 to 70°C (-4 to 158 °F)
T
emperature effects are included in the accuracy.
baud rate 1200, 2400, 4800, 9600, 19200 and
38400. Stop bits 1 or 2, status ON/OFF
Manual keyboard operation
1011 records into a maximum of 20 files,
64k memory
Backlit LCD, 16 character x 2 line alpha numeric
5 to 95% Relative Humidity non condensing
Increase or decrease by a factor of 10
6V Ni-MH rechargeable battery pack (see
warning on page iii)/ mains operation via charger
(supplied).
Regulated DC supply 10V ± 0.1%
max current 10mA
Unregulated DC supply 24V ± 10%
max current 50mA
2