2.8 Date & Time
To select Date & Time proceed as follows:
•
Switch on PC6-IDOS.
•
Press the
key twice to select Set menu.
MENU
•
Select 'Date & Time' option using the
then press the
ENTER
ZERO
2.8.1 To set the time:
•
Select 'Set' using the
•
Select the digit you want to change using the
press
key. Use the
ENTER
ZERO
value and then press
•
Repeat the above step to change the other digits to set
time and proceed to the next set of digits to change the
date.
•
Press the
key to return to the 'Date & Time' menu.
MENU
2.8.2 To view date & time:
•
Select 'View' using the
key.
•
Press the
key to return to the Date & Time menu.
MENU
K0354, Iss. 1
key and
key.
key, then press the
keys to change the
key.
ENTER
ZERO
key and then press the
2.9 Language
•
Press the
•
Scroll to 'Language' using either arrow key and press
key to select.
ENTER
ZERO
•
Select the desired language from the options provided
and press the
Note
: To download different languages, see 3.2 on page 35.
ENTER
key.
ZERO
2.10 Zero (for absolute transducers, please see page 31)
Note: Unlike the 'Tare' function (page 15), 'Zero' reset is a
permanent setting.
To reset pressure reading to zero:
key and
•
Press the
•
Select 'Zero' option using the
key.
ENTER
ZERO
•
Press the
The Pressure Display will now show the pressure value as
zero. If however the pressure reading is not within 5% Full
ENTER
ZERO
Scale the following message will be displayed and the zero
reset will not be possible.
key twice.
MENU
MENU
key.
ENTER
ZERO
key twice to select the Set menu.
MENU
key and then press the
ENTER
key as instructed.
ZERO
Page
30