Raise the Acoustic Output only after attempting image optimization with controls that have no effect on Acoustic Output, such as Gain and
TGC.
NOTE: Refer to the Optimization sections of the Modes chapter for a complete discussion of each control.
Be sure to have read and understood control explanations for each Mode used before attempting to adjust the Acoustic Output control or
WARNING
any control that can effect Acoustic Output.
Acoustic
Use the minimum necessary acoustic output to get the best diagnostic image or measurement during an examination. Begin the exam with
Output
the probe that provides an optimum focal depth and penetration.
Hazard
Acoustic Output Default Levels
In order to assure that an exam may not start at a high output level, the LOGIQ 7 may initiate scanning at a reduced default output level. This reduced level is preset
programmable and depends upon the exam category and probe selected. It takes effect when the system is powered on or New Patient is selected.