LOGIQ 7 Quick Guide
Warning Label Locations
Console Labels
(2)
(7)
(9)
(10)
100-120VÅ`
(8)
220-240VÅ`
Safety
Direction 2291859-100 Rev. 1
(1)
1. Possible shock hazard. Do not remove covers or panels. No user serviceable parts are inside. Refer
servicing to qualified service personnel.
(3)
2. Do not use the following devices near this equipment: cellular phone, radio receiver, mobile radio
transimitter, radio controlled toy, etc. Use of these devices near this equipment could cause this equipment
to perform outside the published specifications. Keep power to these devices turned off when near this
equipment.
(4)
3. The equipment weighs approximately 225 kg (496 lbs). To avoid possible injury and equipment damage
when transporting from one area of use to another:
- Be sure the pathway is clear
- Limit movement to a slow careful walk.
(5)
- Use two or more persons to move the equipment on inclines or long distance.
4. Prescription Device (For U.S.A. Only)
5. The CE Mark of Conformity indicates this equipment conforms with the Council Directive 93/42/EEC.
(6)
6. CISPR
CAUTION: The LOGIQ 7 conforms to the CISPR11, Group 1, Class A of the international standard for
Electromagnetic disturbance characteristics.
7. Voltage Range (Indication label)
8. Signal ground point label
CAUTION: This is only for "FUNCTIONAL GROUNDING", NOT "PROTECTIVE EARTH".
Japan/USA/Asia Console(100V)
9. Power (Indication label)
10. Grounding reliability can only be achieved when this equipment is connected to a receptacle marked
"Hospital Only" or "Hospital Grade".(For U.S.A., Canada, Japan)
Figure 1-2. Label Location (a)
Table 1-8: Label Location Explanations
51
Need help?