Defining the System Setup: Master's Step Setup (Option–Japan Only)
Master's Step Setup (Option–Japan Only)
Select this function to define the parameters for the
Master's Step option.
System Setup
Master's Step
168A,
Figure 12-20. Master's Step
Item
Description
Number of
The number of steps required during the
Steps
exercise portion of the test. This is
calculated from the patient weight, sex and
age, but can be changed here.
Test Type
Test length. Select Single for 1.5 minute
test, Double for 3 minute test or Triple for a
4.5 minute test.
12-34
Figure 12-20. Master's Step (Continued)
Item
Description
Post J(ms)
Number of ms after J point. Used to
determine the ST level
Step Counter
Select Up to display steps taken so far.
Display
Select Down to display steps to go during
exercise.
Continuous
Print rhythm between post exercise ECGs.
Recording
Sound Option
Sets the sound option so that you can
select different tones. The options are
Option 1 , Option 2 , Option 3 , and Off . The
pace of the tone is set by the patient's
weight, sex, and age.
Post Exercise
The time, in minutes, after the 1st post
ECG Time
exercise ECG when a additional ECG
should be taken (up to 9 are available).
Set any undesired tests to 0.
MAC 5000 System — 2000657-024, Revision A