GE MAC 5000 Operator's Manual page 202

Resting ecg analysis system
Hide thumbs Also See for MAC 5000:
Table of Contents

Advertisement

Technical Specifications:
Figure D-4. Safety (Continued)
Item
Patient Leakage Current
Degree of protection against electric
shock
Maintenance Frequency
Repair Guidelines
D-6
Description
<10 µA
Type BF defibrillation protection for the patient cable (acquisition module)
Daily visual inspection and routine cleaning (if needed) performed by user. Use a
commercially available, industrial strength disinfectant cleaner on any part of the
equipment (other than electrodes) which comes into direct contact with the patient.
Every six months routine maintenance checks and test procedures performed by
qualified technical personnel (See the field service manual for details.)
Calibration instructions, equipment descriptions, and all other information which will
assist qualified technical personnel in repairing those parts of the equipment designated
as repairable is available in the field service manual for the equipment.
GE Medical Systems Information Technologies will make available upon request circuit
diagrams and component parts lists for printed circuit boards deemed repairable by
qualified technical personnel.
MAC 5000 System — 2000657-024, Revision A