Master Alarm Setup Screen
Audio Alarms
Re-Alarm
Alarm Limits
Volume
5-18
Setup Procedures: Master Alarm Setup Screen
ALARMS
Select the
softkey to display the Master Alarm Setup screen
Each of the fields on this screen can be accessed under the individual parameter
setup screens. For available settings, refer to
FECG
US
165 172
MASTER ALARM SETUP
AUDIO ALARMS
NBP: ON
MHR/P: ON
RE-ALARM: 120 SEC
ALARM
HIGH
SYSTOLIC:
160
DIASTOLIC:
90
MAP:
140
MHR/P:
120
MSpO
:
100
2
VOLUME: 5
Figure 5-8. Master Maternal Alarm Setup Screen
Individual fields enable/disable the audio alarm functions for NBP, MSpO
MECG. The settings are summarized as follows:
Off: Only a visual indication is provided for an alarm.
On: Both visual and audible (if enabled) indications are provided for an alarm
condition.
When an audio alarm is cancelled using the
indication is removed. However, for MECG and MSpO
paper-load–error condition, an alarm will be re-issued if the alarm state continues
after a specified amount of time. This field adjusts the temporary silence period.
These fields adjust the high and low alarm limits for NBP, MHR/P, and MSpO
This field controls the volume of all alarms.
120 Series Maternal/Fetal Monitor
2015590-001
Appendix A, "Factory Defaults"
TOCO
30
+1
MSpO
: ON
2
LOW
90
MMHG
50
MMHG
50
MMHG
50
BPM
95
%
EXIT
Alarm Silence
pushbutton, the audible
monitoring and during a
2
(Figure
5-8).
.
, and
2
.
2
Revision B