Figure 6. Torque Values for the Flange and Flange Adapters Bolts
Bolt Material
Carbon Steel (CS)
Stainless Steel (SST)

O-rings with Flange Adapters

Use only the O-rings included with the flange adapter for the 3051S ERS sensor. Failure to
install proper fitting flange adapter O-rings may cause process leaks, which can result in
death or serious injury.
When removing flanges or adapters, visually inspect the PTFE O-rings. Replace them if there
are any signs of damage such as nicks or cuts. If replacing O-rings, re-torque the flange bolts
after installation to compensate for seating of the PTFE O-ring.
...
Head Markings
B8M
316
STM
316
316
R
WARNING
Rosemount 3051S ERS System
Bolt
Sensor Module
Isolator Plate
Coplanar Flange
Flange Adapters
B7M
316
SW
316
Initial Torque Final Torque
300 in.-lbs.
650 in.-lbs.
150 in.-lbs.
300 in.-lbs.
7