dIreCT CurrenT ModalITy InforMaTIon

Direct Current Warnings

It is important that excellent coupling be maintained any time current is flowing. Increased resistance from poor coupling or
excessive current density can cause skin reactions.
Electrodes must be attached and probe placed in contact with the patient's skin prior to starting the treatment.
Direct Current tends to lower surface sensory perception as the treatment proceeds. The patient must have good
sensory perception if Direct Current is to be used.
Direct Current Default Setting
Pulse Duration: ................................................................................... 100 mSec
Polarity: ..................................................................................................Negative
Intensity: .............................................................................................................1
The following default settings are set by the manufacturer:
Waveform: .................................................................................................Square
Pulse Duration: .............................................................................. .1-500 mSec
Pulse Duration Selections: .................................. 0.1, 0.3, 0.5, 1, 10, 100, 500
Power Output: ........................................................................................20 mA.
Note: Load impedance does not affect output of the DC waveform.
DYNATRON SOLARIS
PLUS SERIES | SERVICE MANUAL REV. 6
®
| AUGUST 29, 2017
62