Suggested Maintenance Schedule - Dynatronics dynatron solaris 709 plus Service Manual

Table of Contents

Advertisement

general sPeCIfICaTIons
Tri-Wave Light Probe and Pads
DO NOT USE ISOPROPYL ALCOHOL WHEN CLEANING THE TRI-WAVE LIGHT PAD OR PROBES. To
cleanse the probe and pad, use a mild (non caustic) antibacterial soft soap and lukewarm water on a soft cloth. Never
apply soap and water directly on the lens. Gently rinse using the soft, cloth damped with lukewarm water. Dry
by blotting with a damp cloth or chamois. If seeking a commercial cleanser or disinfectant it is recommended that
a product be a disinfectant wipe containing only the following active ingredients:
OctylDecyl Dimethyl Ammonium Chloride
Dioctyl Dimenthyl Ammonium Chloride
Dimenthyl Benzyl Ammonium Chloride
Ultrasound Head
Ultrasound heads should be cleaned with warm water. Always keep the head free from gel buildup. Alcohol may be
used to sterilize the soundhead.
Do not use ice water for cooling soundheads. Do not allow soundheads to overheat repeatedly. This could result in
thermal shock to the crystal. Damage of this type is not covered by the warranty.
ThermoStim Probe
Clean the ThermoStim Probe after each use.
Using a soft cloth, remove all gel residue from the probe.
Using a cotton-tipped applicator, clean the air vents carefully to prevent clogging and to ensure the proper operation
of the probe. Pay particular attention to the vent openings.

Suggested Maintenance Schedule

Service To Be Performed By A Technician:
Every 6 Months
Test leads and carbon electrodes. Lead resistance should be less than 10% above the mean cable resistance. Greater
values indicate strand breakage and lead should be replaced.
Annually
Annual Ultrasound calibration should be performed by a qualified technician.
Check the output voltages and currents on all outputs.
Inspect soundhead connectors on unit and on soundhead.
Verify DAC calibration and current limits.
It is recommended that the Solaris Plus device be sent to the manufacturer for annual calibration including the
Solaris Plus Tri-Wave Light Therapy Probes.
DYNATRON SOLARIS
PLUS SERIES | SERVICE MANUAL REV. 6
®
Didecyl Dimenthyl Ammonium Chloride
Alkyl (C14 50%; C12 40%; C16 10
| AUGUST 29, 2017
122