fInal qualITy CheCK (qC) CheCKoff sheeT
Refer to Test equipment operation manual and unit operating instructions/ manuals
TRI-WAVE LIGHT MEZZANINE OUTPUT:
a)
Verify treatment selections work
b)
Verify light pad operation
c)
Verify light probe operation
9
Refer to Table for probe and pad treatment times
Labels:
10
a)
Apply Serial Number Label
Earth/Enclosure/Patient Leakages(Enclosure/Patient: tester clip to surface of enclosure or Patient lead )
NORMAL (no fault condition + neutral switch closed)
Earth Leakage: Normal/Reversed Polarity, Earth switch depressed (<500 uA)
Enclosure Leakage: Normal/Reversed Polarity, Chassis switch depressed(<100uA)
Patient Leakage: Normal/Reversed Polarity, Chassis switch depressed(<100uA)
SINGLE FAULT (fault condition/ neutral switch open)
Earth Leakage: Normal/Reversed Polarity, Earth switch depressed(<1000uA)
Enclosure Leakage: Normal/Reversed Polarity, Chassis switch depressed(<500uA)
Patient Leakage: Normal/Reversed Polarity, Chassis switch depressed(<500uA)
ACCEPTANCE:
QC, Assembly, Calibration, and Test
D719 Check Sheet
DYNATRON SOLARIS
PLUS SERIES | SERVICE MANUAL REV. 6
®
Wave length
DCP3
Combination
6 Joules
IR / Red / Blue
1:00 min
QC Inspector
DLP3
10 Joules
8:42 min
Initials:
Technician
-
2 of 9-
| AUGUST 29, 2017
Device Specific Procedure
REPAIR/RS
PROD
/
/
/
/
/
/
Date:
DS0408A – REV 1
148