ООО "Техэнком"
Контрольно-измерительные приборы и оборудование
Chapter 11 — Fixed WiMAX Commands
11-9
:TRACe Subsystem
This subsystem contains commands related to the transfer of trace data to and from the
instrument.
:TRACe:PREamble? <trace type>
Parameter: <trace type>
Related Command: :TRACe:DATA?
MS2720T PM
Title: Returns trace header information for the specified trace. The <trace
type> must be one of the following:
SPECtrum|PVTime|ACPR|CONStln|SFLatness|EVSCarrier|EVSY
mbol
Data can be transferred to and from the 7 available display trace types.
Use the commands in the MMEMory subsystem to store and recall
traces from the instrument memory. The response begins with an
ASCII header. The header specifies the number of following bytes. It
looks like #AX, where A is the number of digits in X and X is the
number of bytes that follow the header. Parameters are returned in
comma-delimited ASCII format. Each parameter is returned as
"NAME=VALUE[ UNITS]," Valid parameter names are shown in
Chapter 2.
PN: 10580-00341 Rev. C
www.tehencom.com
11-9 :TRACe Subsystem
11-21
Need help?