Chapter 14 — TD-LTE Commands
:READ:OTA:SCANner?
Description: Triggers a new OTA Scanner measurement and returns the results. The
Default Unit: dBm
Front Panel
Access: Measurements, Over-the-Air, Scanner
:READ:OTA:TXTEst?
Description: Triggers a new OTA Tx Test measurement and returns the results. The
Default Unit: dBm
Front Panel
Access:
MS2720T PM
Title: Read OTA Scanner
OTA Scanner measurement must be the active measurement (specified
by :CONFigure:OTA SCANner. The current measurement can be
queried using the command :CONFigure?
Data is returned as 6 sets of comma-delimited values. Each set as
follows: Cell ID, Group ID, Sector ID, S-SS Power, RSRP, RSRQ, and
SINR
Following the six sets of data is a final value for the Dominance.
"--" is returned for each data that is not valid at that instance.
Title: Read OTA Tx Test
OTA Tx Test measurement must be the active measurement (specified
by :CONFigure:OTA Tx Test. The current measurement can be queried
using the command :CONFigure?
Data is returned as 3 sets of comma-delimited values. Each set as
follows: Cell ID, Group ID, Sector ID, S-SS Power, RSRP, RSRQ, and
SINR
Following the six sets of data is and addition set of four numbers:
Dominance, Antenna count, Average Power, Delta Power.
"--" is returned for each data that is not valid at that instance.