Optional Advanced Seismic Securement - Keithley S530 Administrative Manual

Parametric test system
Hide thumbs Also See for S530:
Table of Contents

Advertisement

S530 Parametric Test System Administrative Guide

Optional advanced seismic securement

If you ordered a system with the advanced seismic fastening option, follow these mounting
instructions.
To install advanced seismic securement:
You must supply the washers (flat and lock), threaded insert anchors, and mounting bolts that attach
the mounting brackets to the floor (16 of each). The bolts should be at least grade-five bolts.
1. Position the cabinet in the desired location on a smooth, level floor.
2. Place the floor mounting brackets at the corners and make sure you have enough room for
proper placement.
3. To properly place the mounting brackets, lift the cabinet by the leveling legs.
Lifting the cabinet will allow the holes in the mounting brackets to line up with the holes in the plinth.
4. Temporarily attach the mounting brackets to each corner with the provided screws.
The following figure shows the restraint brackets and bolt installation for the system cabinet.
S530-924-01 Rev. D / September 2017
Figure 33: S530 seismic restraints
Section 2: S530 site preparation and installation
2-25