3.7 Instrument info
Basic menu ➔
Instrument info
Status message
GOLDCAP WAS EMPTY
WRONG VERSION NO.!
The instrument info window displays general information about the instrument.
It also includes errors "Battery empty" and "Data lost". If one of these instru-
ment errors is present, the info symbol flashes in the status line.
The instrument info is called up from the basic menu:
h Select operating level Instrument info
h Confirm selection with E
3 Operation and visualization
Description
This message appears when a capacitor is built
into the recorder for memory buffering and the
instrument has remained switched off for such a
long time that the capacitor has become
discharged.
Caution:
This will falsify the measurement data.
An attempt was made to read in a configuration
from diskette, but the version numbers of
instrument software and configuration are
different.
Remedy:
Convert configuration diskette via the PC setup
program and create a new one.
41