Move the roller brackets on both sides to the middle position and
secure them firmly.
Position the stainless steel plate over the openings in the water
supply pipe so that one of the outer openings is covered. Secure
the stainless steel plate at the top or bottom with 2 screws. Place
the water connector in the middle opening of the stainless steel
plate so that the outer opening is covered. Secure the water
connector with 4 screws.
Move the roller brackets on both sides to the top position and
secure them firmly.
Position the stainless steel plate over the openings in the water
supply pipe so that the lower opening is covered. Secure the
stainless steel plate at the bottom with 2 screws. Place the water
connector in the upper opening of the stainless steel plate so that
the middle opening is covered. Secure the water connector with 4
screws.
Replace the upper basket on the rails and push it in carefully to
check that the water connection is positioned correctly.