Pressurized DI
water connection
( 30-1,000 kPa) -
depending on
version
Non-pressurized
(gravity feed) DI
water connection
(8.5-60 kPa) -
depending on
version
Demineralized
water ring line
Depending on the version, the machine can be connected to
pressurized DI water with a pressure between 30-1,000 kPa
(4.4-145 psi). If the water pressure is below 200 kPa (29 psi) the water
intake time will be automatically increased.
The pressure tested hose for DI water, coded green, has a ¾ inch
female connection with garden hose thread and is connected to the
onsite DI water tap.
 If the machine is not going to be connected to DI water, the DI
water connection has to be deactivated by a Miele Service
technician. The inlet hose remains in position at the back of the
machine.
If DI water is supplied from a DI water reservoir with gravity feed or
low pressure, a machine version with DI booster pump is required.
The pressure for gravity feed or low pressure ranges between 8.5-60
kPa (1.3 - 8.7 psi). The machine can be converted from pressurized to
non-pressurized or vice-versa. The conversion must only be carried
out by Miele Service.
For non-pressurized DI water reservoirs the connection point to the
machine must be at least as high as the top of the machine. See
installation instructions.
The machine can be connected to a ring line system for
demineralized water. For this purpose, it must be technically adapted
and the controls reset by Miele Professional Service.
Please contact Miele Professional Service for further information.
Water connection
125