6.2.2 Cleaning the Cooling Unit and the Filling Chamber Screen
X Use Cleaning Brush and cotton swabs and paper or cotton cloths for
manual cleaning.
X Use warm rinsing water or isopropyl alcohol as a solvent.
X Clean the Filling Chamber Screen and the Screen of the Cooling Unit with
the Cleaning Brush after each vaporization.
6.2.2.1 Cleaning the Cooling Unit and the Filling Chamber Screen with
rinsing Water
X Place all parts in warm rinsing water for at least 5 minutes.
X Rinse all parts under running water for at least 1 minute.
X If necessary, clean the Cap Lock with warm rinsing water.
X Let all parts dry completely.
6.2.2.2 Cleaning the Cooling Unit and the Filling Chamber Screen with
Isopropyl Alcohol
CAUTION
Isopropyl alcohol is flammable and may easily ignite
Caution! Hot surface!
• Follow the safety instructions of the isopropyl alcohol manufacturer!
NOTICE
Material damage caused by improper cleaning
Bleaching of the Cap Lock Sticker.
Dissolving the adhesive of the Cap Lock Sticker.
• Wipe off the Cap Lock with Signet hologram only with
isopropyl alcohol.
NOTICE
Material damage caused by improper cleaning
Embrittlement of plastic parts.
• Do not immerse plastic parts in isopropyl alcohol for more than
30 minutes.
6 C L E A N I N G / H Y G I E N E
26